More types of mobile phones, and the structure can be divided according to speed and low-speed high-speed mobile phones, high-speed mobile dental handpiece using high pressure air driven turbine handpiece for micro turbine speed can be up to 300,000 to 500,000 r / min, bur cutting speed can be up to 15m / s, while the low-speed mobile dental handpiece because it is gear, speed can be transferred per minute for thousands of 30000 rpm.
2013年8月26日星期一
treatment units Common Troubleshooting
Phone leaks first look is that handsets leak
First, a cell phone leaks
1, if it is leaking tail mobile phone connectors should first check there is no tightening, tighten up if you feel we should look at the end of the phone pad is intact. Tail pad bad need of replacement. If you have not resolved the issue will look at the inner core is intact cell phone connector.
2, mobile phone front air dental lab control valve leaking first look at whether the work intact. The gas control valve outlet pipe pull out, stepping on the foot switch, and release water out there that air control valve problems. Workaround, open the air control valve spool look out O-ring is damaged, if not broken out to put a spring elongated look, and then the O-ring position, cast a little Vaseline, be careful not wipe too much. Mounted on top. If installed after leaking or listen stepped foot there deflated after releasing the sound, if the foot switch deflated normal phone line that is the problem. (Gas pedal and hold this problem rarely occurs) Solution: Put the phone from water and gas mains, disconnect integrated valve, disposable needle with a suction tube connector end of the line, Sheung Shui, recoil from the phone to solve the problem.